Precision of high definition spectral-domain optical coherence tomography for measuring central corneal thickness.

Tipo: Artículo
Código: PMID: 22395881
Autores: Correa-Pérez ME, López-Miguel A, Miranda-Anta S, Iglesias-Cortiñas D, Alió JL, Maldonado MJ.
Títuto Revista: Invest Ophthalmol Vis Sci.
Referencia:
Colaboradores: Instituto Universitario de OftalmoBiología Aplicda (IOBA), Universidad de Valladolid (Rd07-0062-0013); Instituto Oftalmológico de Alicante (IOA), Facultad de Medicina, Universidad Miguel Hernández (RD07-0062-0018).
Centro: G05-Barcelona

Abstract

PURPOSE:

This study was intended to assess the reliability of central corneal thickness (CCT) measurements using Cirrus high-definition optical coherence tomography (HD-OCT) in healthy subjects and its accuracy compared with ultrasonic pachymetry.

METHODS:

Seventy-seven consecutive subjects were recruited for evaluating repeatability, and agreement between two examiners. To analyze repeatability, one examiner measured 77 eyes four times in succession. To study agreement between two observers, a second independently trained examiner obtained another CCT measurement. We also measured eyes in a subgroup of 20 patients using standard ultrasonic pachymetry. Within-subject standard deviation (S(w)), coefficient of variation (CV), limits of agreement (LoA), and intraclass correlation coefficient (ICC) data were obtained.

RESULTS:

For repeatability, the S(w) and precision (1.96 × S(w)) were 4.86 and 9.52 μm, respectively. Intraobserver CV was 0.89% and the ICC was 0.98 (95% confidence interval [CI], 0.97-0.99). For agreement between two examiners, the S(w) and precision were 7.58 and 14.85 μm, respectively; the CV was 1.40%. The mean difference between observers was -0.13 μm (95% CI, -1.85 to 1.58; P = 0.87). The width of the LoA was 29.64 μm. Median difference between Cirrus HD-OCT and ultrasound CCT measurements was -4.5 μm (interquartile range, -7.0-0.0; P = 0.04).

CONCLUSIONS:

Cirrus HD-OCT provides repeatable CCT measurements, good agreement between two independently trained examiners, and its systematic bias compared to ultrasonic pachymetry is clinically negligible. Therefore, research laboratories and eye clinics using Cirrus HD-OCT as a diagnostic imaging method, can also benefit from a reliable noncontact pachymeter when counseling patients with glaucoma and those undergoing corneal and refractive surgeries.